Abstract Views :151 |
PDF Views:0
Authors
Affiliations
1 College of Pharmacy, Shri R.M.S. Institute of Science and Technology, Bhanpura-458775, Dist. Mandsaur, (M.P.), IN
Source
Asian Journal of Research in Chemistry, Vol 3, No 3 (2010), Pagination: 604-607
Abstract
A precise and accurate reverse phase high-performance liquid chromatographic (RP-HPLC) method was developed and validated for estimation of quetiapine fumarate. The mobile phase consisted of methanol:water:acetonitrile (50:40:10) was used at a flow rate of 1 ml/min and pH value of 7.0, adjusted with orthophosphosphoric acid. The detection was carried out at a wavelength of 251 nm. The developed method was found to be in accordance with ICH guidelines with respect to accuracy, precision, reproducibility, linearity and specificity.
Keywords
Quetiapine Fumarate, RP-HPLC, Validation.