The adsorption of carbon dioxide (CO2) on solid sorbents or any gas adsorption in general can be experimentally determined using various techniques. The present study demonstrates a dynamic column breakthrough method for the measurement of CO2 adsorption using a cost-effective purpose-built single-column fixed-bed adsorption unit. The CO2 adsorption capacity is estimated using a graphical method that does not require knowledge of the variation of molar CO2 flow rate at the column exit. The experimental method is validated using a commercial zeolite 13X adsorbent.

Keywords

Carbon Dioxide Adsorption, Dynamic Column Breakthrough Method, Gas Chromatography, Zeolite and Sorbent.
User
Notifications
Font Size

Abstract Views: 220

PDF Views: 78




  • Measurement of CO2 Adsorption Using the Cost-Effective Dynamic Column Breakthrough Method

Abstract Views: 220  |  PDF Views: 78

Authors

Lakshminarayana Kudinalli Gopalakrishna Bhatta
Centre for Incubation, Innovation, Research and Consultancy, Jyothy Institute of Technology, Tataguni, Off Kanakapura Road, Bengaluru 560 082, India
Seetharamu Subramanyam
R&D Centre, BMS College of Engineering, P.O. Box No. 1908, Bull Temple Road, Bengaluru 560 019, India
Madhusoodana D. Chengala
Ceramic Technological Institute, Bharat Heavy Electricals Ltd, Bengaluru 560 012, India
Umananda Manjunatha Bhatta
Centre for Incubation, Innovation, Research and Consultancy, Jyothy Institute of Technology, Tataguni, Off Kanakapura Road, Bengaluru 560 082, India
Krishna Venkatesh
Centre for Incubation, Innovation, Research and Consultancy, Jyothy Institute of Technology, Tataguni, Off Kanakapura Road, Bengaluru 560 082, India
V. Raghavendra
Mayura Analytical LLP, #455, 18th Main, 4th T Block, Jayanagar, Bengaluru 560 041, India

Abstract


The adsorption of carbon dioxide (CO2) on solid sorbents or any gas adsorption in general can be experimentally determined using various techniques. The present study demonstrates a dynamic column breakthrough method for the measurement of CO2 adsorption using a cost-effective purpose-built single-column fixed-bed adsorption unit. The CO2 adsorption capacity is estimated using a graphical method that does not require knowledge of the variation of molar CO2 flow rate at the column exit. The experimental method is validated using a commercial zeolite 13X adsorbent.

Keywords


Carbon Dioxide Adsorption, Dynamic Column Breakthrough Method, Gas Chromatography, Zeolite and Sorbent.



DOI: https://doi.org/10.18520/cs%2Fv112%2Fi04%2F835-838